Data Foundry

Medical device recalls 2015

Ysio Max — Class II medical device recall Z-0994-2015

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2015-01-28, distributed in ID, IL, MI, MN, MO, ND, PA. Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 an

Recall numberZ-0994-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2014-12-15
Classified2015-01-16
Reported by FDA2015-01-28
Terminated2017-01-26
DistributedID, IL, MI, MN, MO, ND, PA
Quantity11
Reason classeslabeling, software
Serial numbers24015, 24026, 24038, 24039, 24051, 24058, 24059, 24066, 24075, 24077, 24090

Product

Ysio Max, Luminos dRF Max and Agile Max systems with software version VE10 and SmartOrtho license. The Ysio Max is a radiographic system used in hospitals, clinics, and medical practices. Ysio Max enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio Max system is not meant for mammography. The Ysio Max uses integrated or portable digital detectors for generating diagnostic images by converting x-rays into electronic signals. Ysio Max is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

Reason for recall

Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.