Data Foundry

Medical device recalls 2023

Alinity m System — Class II medical device recall Z-1009-2023

Ongoing recall by Abbott Molecular, Inc., reported 2023-02-01. The firm identified the following three potential issues which may have impacted ADUs installed on distributed

Recall numberZ-1009-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAbbott Molecular, Inc., Des Plaines, IL, United States
Recall initiated2022-12-09
Classified2023-01-20
Reported by FDA2023-02-01
Quantity918 units
Model numbers08N53-002

Product

Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.

Reason for recall

The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1009-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.