The GLOBAL UNITE Platform Shoulder System is intended for cemented or… — Class II medical device recall Z-1012-2018
Terminated recall by DePuy Orthopaedics, Inc., reported 2018-03-21. The screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted durin
| Recall number | Z-1012-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | DePuy Orthopaedics, Inc., Warsaw, IN, United States |
| Recall initiated | 2017-11-28 |
| Classified | 2018-03-14 |
| Reported by FDA | 2018-03-21 |
| Terminated | 2020-03-27 |
| Quantity | 85 |
| Reason classes | device malfunction |
| Lot numbers | 8527674, 8556802, 8556803, 8556804, 8556805, 8556910, 8556951, 8564487, 8572431, 8583381, 8605849, 8610357 |
Product
The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Product labeling: LE¿O-1100-30-000 REV D.pdf.pdf, LE¿O-1100-40-000 REV D.pdf.pdf, LE¿O-110030100_REV D.pdf.pdf, LE¿O-110030110_REV D.pdf.pdf, LE¿O-110030120_REV D.pdf.pdf
Reason for recall
The screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays.