Data Foundry

Medical device recalls 2013

Microbiologics KWIK-QC SLIDES — Class II medical device recall Z-1021-2013

Terminated recall by Microbiologics Inc, reported 2013-04-03, distributed in CA, IN, MS, NY, OH, PA, SD. Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWI

Recall numberZ-1021-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMicrobiologics Inc, Saint Cloud, MN, United States
Recall initiated2013-02-11
Classified2013-03-28
Reported by FDA2013-04-03
Terminated2016-03-10
DistributedCA, IN, MS, NY, OH, PA, SD
Countries outside the USAE, AT, AU, BE, CA, CO, JP, MY, TR, TW, VE, ZA
Reason classesmicrobial contamination
Lot numbers4239, 4243, 4333, 4337, SL43-10

Product

Microbiologics KWIK-QC SLIDES ( KWIK-QC Acid Fast Stain Slides and KWIK-QC Mycobacterium Slides). Slide preparations serving as quality control challenges to demonstrate satisfactory performance of staining reagents.

Reason for recall

Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWIK-QC Mycobacterium, SL43-10. These products are being recalled because the control organisms are not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could pote