CyberKnife Treatment Delivery System — Class II medical device recall Z-1023-2025
Completed recall by Accuray Incorporated, reported 2025-02-05, distributed nationwide. Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrol
| Recall number | Z-1023-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Completed |
| Voluntary or mandated | Voluntary |
| Recalling firm | Accuray Incorporated, Madison, WI, United States |
| Recall initiated | 2024-12-23 |
| Classified | 2025-01-27 |
| Reported by FDA | 2025-02-05 |
| Distributed | Nationwide (US) |
| Quantity | 212 |
| Reason classes | device malfunction |
| Model numbers | 0660000 |
Product
CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.
Reason for recall
Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.