Data Foundry

Medical device recalls 2025

CyberKnife Treatment Delivery System — Class II medical device recall Z-1023-2025

Completed recall by Accuray Incorporated, reported 2025-02-05, distributed nationwide. Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrol

Recall numberZ-1023-2025
Product typeMedical device
ClassificationClass II
StatusCompleted
Voluntary or mandatedVoluntary
Recalling firmAccuray Incorporated, Madison, WI, United States
Recall initiated2024-12-23
Classified2025-01-27
Reported by FDA2025-02-05
DistributedNationwide (US)
Quantity212
Reason classesdevice malfunction
Model numbers0660000

Product

CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Reason for recall

Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.