Medtronic Prostiva RF Therapy Generator — Class II medical device recall Z-1031-2013
Terminated recall by Medtronic Neuromodulation, reported 2013-04-10, distributed nationwide. Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930
| Recall number | Z-1031-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Neuromodulation, Minneapolis, MN, United States |
| Recall initiated | 2013-02-27 |
| Classified | 2013-04-02 |
| Reported by FDA | 2013-04-10 |
| Terminated | 2013-11-21 |
| Distributed | Nationwide (US) |
| Model numbers | 8930 |
Product
Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.
Reason for recall
Medtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi