Arial 900 MHz Call Station — Class II medical device recall Z-1032-2026
Ongoing recall by Securitas Healthcare LLC, reported 2026-01-14, distributed nationwide. Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may c
| Recall number | Z-1032-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Securitas Healthcare LLC, Lincoln, NE, United States |
| Recall initiated | 2025-12-08 |
| Classified | 2026-01-08 |
| Reported by FDA | 2026-01-14 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA |
| Quantity | 1,898 units |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00840948102660 |
| Serial numbers | 000CCC0E45AF, 000CCC0E45B0, 000CCC0E45B1, 000CCC0E45B2, 000CCC0E45B3, 000CCC0E45B4, 000CCC0E45B5, 000CCC0E45B6, 000CCC0E45B7, 000CCC0E45B8, 000CCC0E45B9, 000CCC0E45BA, 000CCC0E45BB, 000CCC0E45BC, 000CCC0E45BD, 000CCC0E45BE, 000CCC0E45BF, 000CCC0E45C0, 000CCC0E45C1, 000CCC0E45C2, 000CCC0E45C3, 000CCC0E45C4, 000CCC0E45C5, 000CCC0E45C6, 000CCC0E45D3 and 475 more |
| Model numbers | CSK200-900MR |
Product
Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3
Reason for recall
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1032-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.