Data Foundry

Medical device recalls 2019

PET Plates Coated Film-MONOMEDBV2S2- Radiographic Film REF: XJUB8 — Class III medical device recall Z-1034-2019

Terminated recall by Agfa N.V., reported 2019-03-27, distributed nationwide. Due to an inhomogeneous coating solution, pinholes can become visible in even dark images. Visibility of the p

Recall numberZ-1034-2019
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAgfa N.V., Mortsel, N/A, Belgium
Recall initiated2019-01-25
Classified2019-03-21
Reported by FDA2019-03-27
Terminated2019-06-14
DistributedNationwide (US)
Quantity2,029 boxes
Lot numbers36930048, XJM73, XJUB8
Model numbersXJM73, XJUB8

Product

PET Plates Coated Film-MONOMEDBV2S2- Radiographic Film REF: XJUB8, XJM73, XJUDD, XJM85, EKLYZ. Product Usage: Due to an inhomogeneous coating solution, pinholes can become visible in even dark images. Visibility of the pinholes varies as material from this coating run shows pinholes in different frequencies and intensities.

Reason for recall

Due to an inhomogeneous coating solution, pinholes can become visible in even dark images. Visibility of the pinholes varies as material from this coating run shows pinholes in different frequencies and intensities

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1034-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.