CartoUnivu Module — Class II medical device recall Z-1045-2014
Terminated recall by Biosense Webster, Inc., reported 2014-02-26, distributed in CA, KY, MA, NJ, UT. Biosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3)
| Recall number | Z-1045-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Biosense Webster, Inc., Irwindale, CA, United States |
| Recall initiated | 2014-02-12 |
| Classified | 2014-02-20 |
| Reported by FDA | 2014-02-26 |
| Terminated | 2014-05-23 |
| Distributed | CA, KY, MA, NJ, UT |
| Countries outside the US | AT, BE, DE |
| Serial numbers | 11115, 11125, 11320, 11519, 13201 |
| Model numbers | KT-5400-124 |
Product
CartoUnivu Module, Model: Carto 3 Software versions 3.2.2 and 3.2.3; Catalog No. KT-5400-124 WITH THE CARTOUNIVU MODULE, FLUOROSCOPIC CAPTURES (IMAGES AND CINE CLIPS) CAN BE IMPORTED TO AND DISPLAYED IN THE CART0 3 SYSTEM. THESE IMAGES ARE NOT FOR DIAGNOSTIC PURPOSES. WHEN FLUOROSCOPIC CAPTURES ARE INTEGRATED INTO THE CART0 3 SYSTEM MAP VIEWERS (MAP VIEWER AND ADDITIONAL VIEW), CART0 3 MAPS, MRl OR CT IMAGES, AND CATHETER VISUALIZATION ARE SUPERIMPOSED ON THE FLUOROSCOPIC CAPTURE.
Reason for recall
Biosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only. If the table rotation is not returned to zero point, there is a misalignment between the Carto 3 System map display and the fluoroscopic capture.