Data Foundry

Medical device recalls 2013

Siemens Dimension — Class III medical device recall Z-1046-2013

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2013-04-10. Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in Syst

Recall numberZ-1046-2013
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2013-02-27
Classified2013-04-01
Reported by FDA2013-04-10
Terminated2014-01-14
Quantity2,207 total

Product

Siemens Dimension(R) EXL(TM) LM (Siemens Material Number 10486890). Intended to measure a variety of analytes in human body fluids.

Reason for recall

Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1046-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.