Data Foundry

Medical device recalls 2016

Large Volume Paracentesis Kit is packaged in a thermaformed tray… — Class II medical device recall Z-1047-2016

Terminated recall by Gi Supply, reported 2016-03-16, distributed in CA, DC, FL, IA, KY, MA, MO, NH…. Customers have reported that the pinch clamp on the RP Bag was missing. Ascetic fluid from the abdominal cavit

Recall numberZ-1047-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGi Supply, Camp Hill, PA, United States
Recall initiated2016-01-28
Classified2016-03-04
Reported by FDA2016-03-16
Terminated2016-12-07
DistributedCA, DC, FL, IA, KY, MA, MO, NH, NJ, NY, OH, OR, PA, SC, TN
Quantity390 kits
Lot numbers2018/04, 61442434
Model numbersGIS-28

Product

Large Volume Paracentesis Kit is packaged in a thermaformed tray which contains slots and snap-in holders for most components. Packed sterile in Tyvek bag. Catalog #: GIS-28 Lot #: 61442434 Expiration Date: 2018/04

Reason for recall

Customers have reported that the pinch clamp on the RP Bag was missing. Ascetic fluid from the abdominal cavity may splash User if bag is not clamped prior to removal.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1047-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.