Data Foundry

Medical device recalls 2015

Persona — Class II medical device recall Z-1052-2015

Terminated recall by Zimmer, Inc., reported 2015-02-11, distributed nationwide. Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning

Recall numberZ-1052-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer, Inc., Warsaw, IN, United States
Recall initiated2014-12-11
Classified2015-02-04
Reported by FDA2015-02-11
Terminated2015-10-23
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, CA, CH, CZ, DE, DO, ES, FR, GB, IL, IN, IR, IT, JP, KR, LU, NL, SA, SE, SG, ZA
Quantity67,963
Reason classesdevice malfunction
Model numbers42-5279-002-00

Product

Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The Persona TASPs are sterilizable instruments Intended for multiple uses. The TASP components are used to aid the surgeon in assessing the thickness of the articular surface required. The TASP construct is utilized while conducting a trial range of motion. The TASP consists of 3 parts: a TASP bottom, a TASP shim, and a TASP top. The shims come in varying thicknesses to allow trialing for specific Articular Surface implant thicknesses.

Reason for recall

Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.