MultiDiagnost-Eleva — Class II medical device recall Z-1052-2019

Terminated recall by Philips North America, LLC, reported 2019-04-03, distributed nationwide. The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement tha

Recall numberZ-1052-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, LLC, Andover, MA, United States
Recall initiated2019-02-28
Classified2019-03-23
Reported by FDA2019-04-03
Terminated2020-08-03
DistributedNationwide (US)
Quantity1,969

Product

MultiDiagnost-Eleva, Product codes 708032, 708036 Product Usage: As a multifunctional I universal imaging application system, General RIF, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications on human patients. This includes the following general areas: Digestive system: Swallowing studies, Oesophagus, Stomach, Small intestine, Colon, Defeacography, ERCP, T-tube cholangiogram, Liver biopsies, Transjugular Intrahepatic Portosystemic Shunts (TIPS) Skeletal system: Bone studies. Urinary system: IVP, Cystograms, Percutaneous, Nephrolithotomy, Nephrostomy tube replacement Reproductive system: Hysterosalpingogram, Vena spermatica, Cavernography Respiratory system: Thorax, Bronchoscopy, Pulmonary biopsies Circulatory system: Venography, Arteriography, Thrombolytic Therapy, Em bolizations, Em bolectomy, IVC filter placement, Dilatations, Stent placement. Various: Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.

Reason for recall

The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.