Data Foundry

Medical device recalls 2023

GE Nuclear Medicine 600/800 series systems — Class I medical device recall Z-1056-2023

Ongoing recall by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, reported 2023-02-15, distributed nationwide. A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-thr

Recall numberZ-1056-2023
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, Tirat Carmel, Israel
Recall initiated2022-12-19
Classified2023-02-08
Reported by FDA2023-02-15
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EG, ES, FI, FR, GB, GH, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, LT, MA, MX, MY, NG, NL, NO, NZ, PA, PE, PH, PK, PL, PT, QA, RO, RS, RU, SA, SE, SG, SK, SV, TH, TR, TW, VN, ZA
Quantity34 units
Serial numbers0000064GY51011, 26043, 64GX51001, 64GY51002, 64GY51003, 64GY51004, 64GY51005, 64GY51006, 64GY51007, 64GY51008, 64GY51009, 64GY51010, 64GY51011, 64GY51012, 64GY51013, 64GY51014, 64GY51015, 64GY51016, 64GY51017, 64GY51018, 64GY51019, 64GY51020, 64GY51021, 64GY51022, 64GY51023 and 9 more
Model numbers5376204-41, H3100JZ, H3101RL

Product

GE Nuclear Medicine 600/800 series systems, specifically: a) OPTIMA O640 X 3/8 inch NR, b) DISCOVERY O640 X 3/8 inch, c) NM OPTIMA O640 X 3/8 inch, d) O640 NM Gantry 3/8 inch, e) O640 NM GANTRY FINAL ASSY, f) OPTIMA O640 X 3/8 inch; Emission Computed Tomography System

Reason for recall

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1056-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.