Data Foundry

Medical device recalls 2014

Synergy XVI XVI can incorrectly calculate the target position of the… — Class II medical device recall Z-1059-2014

Terminated recall by Elekta, Inc., reported 2014-02-26, distributed nationwide. XVI can incorrectly calculate the target position of the treatment table.

Recall numberZ-1059-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmElekta, Inc., Atlanta, GA, United States
Recall initiated2013-11-25
Classified2014-02-20
Reported by FDA2014-02-26
Terminated2017-11-24
DistributedNationwide (US)
Countries outside the USAT, AU, BE, BG, BH, BR, CA, CH, CN, CY, CZ, DE, DK, DZ, EC, EE, ES, FI, FR, GB, GR, HK, HR, HU, IE, IL, IN, IQ, IT, JO, JP, KR, MA, MX, MY, NL, NO, NZ, PK, PL, PT, RO, RU, SA, SE, SG, SI, TH, TR, TW, VE, ZA
Quantity1,028
Model numbers4.5, 4.5.1, 4.6, 5.0

Product

Synergy XVI XVI can incorrectly calculate the target position of the treatment table, because of an error that does not occur very frequently. This is the sequence of events when this error occurs - (1) You do an online VolumeView acquisition and registration. (2) You accept the registration and the Table Move Assistant dialog box appears. (3) The Relative Actual column of the Table Move Assistant dialog box incorrectly appears blank. (4) When this occurs, XVI does not read the Absolute Actual position of the treatment table. This causes XVI to calculate the treatment table moves from the zero position of the treatment table (+0.0, + 0.0, + 0.0). (5) In XVI R5.0, if the Relative Set values are less than or equal to + or -2.5cm (for earlier releases of XVI, the value is + or -2cm), you can do remote automatic table movement (RATM) from the function keypad (FKP). (6) When you press the Table ASU (automatic setup) buttons, the treatment table moves to a position that was incorrectly calculated from XVI.

Reason for recall

XVI can incorrectly calculate the target position of the treatment table.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1059-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.