Data Foundry

Medical device recalls 2015

CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon… — Class II medical device recall Z-1062-2015

Terminated recall by Maquet Datascope Corp - Cardiac Assist Division, reported 2015-02-18. Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fibe

Recall numberZ-1062-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMaquet Datascope Corp - Cardiac Assist Division, Mahwah, NJ, United States
Recall initiated2014-12-16
Classified2015-02-06
Reported by FDA2015-02-18
Terminated2017-06-09
Model numbers0998-00-0800-XX

Product

CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.

Reason for recall

Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1062-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.