Data Foundry

Medical device recalls 2025

Intact Vascular Tack Endovascular System — Class I medical device recall Z-1070-2025

Ongoing recall by Spectranetics Corporation, reported 2025-02-26, distributed nationwide. Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by t

Recall numberZ-1070-2025
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSpectranetics Corporation, Colorado Springs, CO, United States
Recall initiated2025-01-10
Classified2025-02-14
Reported by FDA2025-02-26
DistributedNationwide (US)
Quantity2,939
Reason classesspecification failure
UPC / GTIN / UDI-DI00850003494043, 00863328000103, 00863328000189
Lot numbers154150042/328641, 349049
Model numbers154090042, 154090042/322101, 154150041, 154150042, 156080062, 156080062/302251, 156135061, 156135062, 156135062/271330, 206080062, 206080062/321083, 206135061, 206135062, 206135062/328643, 283066, 336121, 336152, 345567, 349034, 349035

Product

Intact Vascular Tack Endovascular System, to treat vascular dissections with Tack implant(s) following angioplasty: (4F,1.5-4.5mm), 150cm, REF: 154150041; Tack, 4F Gen 1.5, 150cm CE, REF: 154150042; Tack, 4F Gen 1.5, 90cm CE, REF: 154090042 (6F, 3.5 - 6.0mm), 135cm, REF: 156135061; (6F, 4.0 - 8.0mm), 135cm, REF: 206135061; Tack, 6F Gen 2.0, 135cm CE, REF: 206135062; Tack, 6F Gen 1.5, 135cm CE, REF: 156135062; Tack, 6F Gen 2.0, 80cm CE, REF: 206080062; Tack, 6F Gen 1.5, 80cm CE, REF: 156080062

Reason for recall

Use of Tack Endovascular system, designed to treat acute dissections of inner wall or lining of an artery by tacking damaged tissue to inner luminal surface, could result in problems with deployment and stability of device in vessel after deployment, which may cause failure to resolve dissection, migration of implant, bailout stenting, reintervention, unintended removal of tack devices, and ischemia.