All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System… — Class I medical device recall Z-1074-2013

Terminated recall by Lifescan Inc, reported 2013-04-24, distributed nationwide. The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying

Recall numberZ-1074-2013
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmLifescan Inc, Milpitas, CA, United States
Recall initiated2013-03-11
Classified2013-04-12
Reported by FDA2013-04-24
Terminated2015-08-03
DistributedNationwide (US)
Quantity710,413

Product

All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch¿ VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be used for testing multiple patients. The OneTouch¿ VerioIQ Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

Reason for recall

The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.