The SurgiWrap Bioresorbable Adhesion Barrier Film — Class II medical device recall Z-1076-2015
Terminated recall by MAST Biosurgery USA, Inc, reported 2015-02-18. The expiration date of the bar code on the envelope label did not match the expiration date next to the hourgl
| Recall number | Z-1076-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | MAST Biosurgery USA, Inc, San Diego, CA, United States |
| Recall initiated | 2014-08-26 |
| Classified | 2015-02-10 |
| Reported by FDA | 2015-02-18 |
| Terminated | 2015-02-24 |
| Countries outside the US | CN |
| Quantity | 49 units |
| Reason classes | labeling |
| UPC / GTIN / UDI-DI | 18964050255027 |
| Lot numbers | 2017-08, 52804, REF25502-01 |
| Model numbers | 25502-01 |
Product
The SurgiWrap Bioresorbable Adhesion Barrier Film. SurgiWrap 100x130x0.05mm (01) 18964050255027 (17) 170731(10)52804 LOT 52804 2017-08 REF25502-01 Product is wrapped in a foil pouch that is then sealed into an additional envelope.
Reason for recall
The expiration date of the bar code on the envelope label did not match the expiration date next to the hourglass. The printed expiration date was August of 2017, while the bar code date read as July 31, 2017. The correct expiration date was July 31, 2017.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1076-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.