Data Foundry

Medical device recalls 2017

Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system — Class II medical device recall Z-1076-2017

Terminated recall by INO Therapeutics (dba Ikaria), reported 2017-02-01, distributed in CA, CO, DC, FL, GA, HI, IA, IL…. Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which

Recall numberZ-1076-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmINO Therapeutics (dba Ikaria), Madison, WI, United States
Recall initiated2016-11-21
Classified2017-01-23
Reported by FDA2017-02-01
Terminated2017-12-20
DistributedCA, CO, DC, FL, GA, HI, IA, IL, IN, MI, MN, NC, NJ, NM, NY, OH, PA, TX
Quantity45
Serial numbersDS20070030, DS20070043, DS20070094, DS20070268, DS20090096, DS20090458, DS20090484, DS20090539, DS20090620, DS20090645, DS20100071, DS20100118, DS20100216, DS20100266, DS20100268, DS20100330, DS20100439, DS20100496, DS20100604, DS20100675, DS20100744, DS20100762, DS20100950, DS20101199, DS20101337 and 20 more
Model numbers10007

Product

Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for inhalation). The INOmax DSIR uses a dual-channel design to ensure the safe delivery of INOMAX. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and 02 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.

Reason for recall

Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate proportional valve calibration table in an affected device.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1076-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.