Data Foundry

Medical device recalls 2025

AlluraXper FD20 Biplane OR Table System Code: — Class II medical device recall Z-1082-2025

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2025-02-12, distributed nationwide. Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Phil

Recall numberZ-1082-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands
Recall initiated2025-01-13
Classified2025-02-04
Reported by FDA2025-02-12
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, KY, LB, LK, LT, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA
Quantity1 units

Product

AlluraXper FD20 Biplane OR Table System Code: (1) 722020 (2) 722025

Reason for recall

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel