Medtronic MiniMed Paradigm Insulin Infusion Pumps Model Numbers:… — Class II medical device recall Z-1085-2013
Terminated recall by Medtronic MiniMed, reported 2013-04-17, distributed nationwide. Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap may become det
| Recall number | Z-1085-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic MiniMed, Northridge, CA, United States |
| Recall initiated | 2013-03-22 |
| Classified | 2013-04-10 |
| Reported by FDA | 2013-04-17 |
| Terminated | 2013-11-04 |
| Distributed | Nationwide (US) |
| Quantity | 428,000 units |
| Reason classes | device malfunction |
| Model numbers | MMT-511, MMT-512, MMT-515, MMT-522, MMT-522K, MMT-523, MMT-523K, MMT-554, MMT-712, MMT-712E, MMT-715, MMT-722, MMT-722K, MMT-723, MMT-723K, MMT-754 |
Product
Medtronic MiniMed Paradigm Insulin Infusion Pumps Model Numbers: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754. Product Usage: Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.
Reason for recall
Medtronic is recalling the Medtronic Paradigm Insulin Pump because the pump's drive support cap may become detached from the pump case and protrude from the lower right side of the pump. When the pump is expose to water it may result in damage to the pump's internal electronics. This moisture damage can prevent the pump's buttons from working properly or can cause the pump to alarm.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1085-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.