Data Foundry

Medical device recalls 2016

Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The… — Class II medical device recall Z-1088-2016

Terminated recall by Medtronic Navigation, Inc., reported 2016-03-16. Screws holding the detector panel in as a result the detector panel could make contact with other parts with i

Recall numberZ-1088-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Navigation, Inc., Littleton, MA, United States
Recall initiated2016-01-29
Classified2016-03-08
Reported by FDA2016-03-16
Terminated2016-03-10
Countries outside the USAT
Quantity1 units
Serial numbersC0936
Model numbersBL-700-02000

Product

Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.

Reason for recall

Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1088-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.