Medline medical procedure convenience kits labeled as: 1) ROBOTIC — Class II medical device recall Z-1088-2026
Ongoing recall by Medline Industries, LP, reported 2026-01-21, distributed nationwide. Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigatio
| Recall number | Z-1088-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medline Industries, LP, Northfield, IL, United States |
| Recall initiated | 2025-12-16 |
| Classified | 2026-01-14 |
| Reported by FDA | 2026-01-21 |
| Distributed | Nationwide (US) |
| Quantity | 12,584 units |
| Reason classes | packaging |
| UPC / GTIN / UDI-DI | 10193489867985, 10193489927863, 10193489956504, 10193489979701, 10195327176426, 10195327209766, 10195327213886, 10195327251291, 10195327261504, 10195327270193, 10195327342593, 10195327343392, 10195327362690, 10195327389789, 10195327391256, 10195327392987, 10195327395896, 10195327409142, 10195327436346, 10195327456764, 10195327474881, 10195327498481, 10195327500993, 10195327504984, 10195327504991 and 167 more |
| Lot numbers | 25BBH373, 25BBJ047, 25BBR098, 25BBT778, 25BBU398, 25BBU420, 25BDB112, 25BDB159, 25BDB494, 25BME871, 25BME966, 25BMF447, 25BMF583, 25BMF759, 25BMI178, 25BMI302, 25BMI579, 25CBA952, 25CBB606, 25CBC157, 25CBC542, 25CBC570, 25CBC628, 25CBD568, 25CBE442, 25CBE740, 25CBH604, 25CBH661, 25CBH758, 25CBI860, 25CBJ023, 25CBJ638, 25CBK354, 25CBK432, 25CBK831, 25CBL682, 25CBL781, 25CBM352, 25CBM414, 25CBN673 and 190 more |
| Model numbers | CDS980523M, CDS980647K, CDS980864Q, CDS981583U, CDS981583V, CDS981826K, CDS982377K, CDS982779N, CDS982780V, CDS982857K, CDS983393L, CDS984656Q, CDS984656R, CDS984663J, CDS985467J, DYNJ36840L, DYNJ38152D, DYNJ50649J, DYNJ52096P, DYNJ52564D, DYNJ52564F, DYNJ52581D, DYNJ53782J, DYNJ53968P, DYNJ57458B and 71 more |
Product
Medline medical procedure convenience kits labeled as: 1) ROBOTIC, REF CDS980523M; 2) ROBOTIC GYN, REF CDS980647K; 3) DAVINCI PROCEDURE, REF CDS980864Q; 4) GYN-URO LAPSCP CDS-LF, REF CDS981583U; 5) GYN-URO LAPSCP CDS-LF, REF CDS981583V; 6) GYN ROBOTIC, REF CDS981826K; 7) DEPAUL DAVINCI CDS, REF CDS982377K; 8) GU DAVINCI, REF CDS982779N; 9) GYN DAVINCI, REF CDS982780V; 10) UNIVERSAL ROBOTIC ORLANDO, REF CDS982857K; 11) ROBOTIC GU/GYN, REF CDS983393L; 12) GYN LAPAROSCOPY CDS, REF CDS984656Q; 13) GYN LAPAROSCOPY CDS, REF CDS984656R; 14) DAVINCI CDS UCC-LF, REF CDS984663J; 15) ROBOTIC UROLOGY PROSTATE, REF CDS985467J; 16) DAVINCI PROSTATECTOMY, REF DYNJG901002B; 17) ROBOTIC PACK IMF 100913- LF, REF DYNJVB91075C; 18) ROBOTIC PACK, REF DYNJ36840L; 19) ROBOTIC PACK MNMC, REF DYNJ38152D; 20) WMC GYN LAPAROSCOPY PACK-LF, REF DYNJ50649J; 21) ROBOTIC GASTRIC BYPASS PACK, REF DYNJ52096P; 22) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564D; 23) SCRIPPS SW ROBOTIC PACK, REF DYNJ52564F; 24) GYN ROBOTIC PACK, REF DYNJ52581D; 25) GYN LAPAROSCOPY PACK, REF DYNJ53782J; 26) UNIVERSAL DAVINCI ROBOTIC P, REF DYNJ53968P; 27) WALKER GYN-UROLOGY ROBOTIC PK, REF DYNJ57458B; 28) GYN LAPAROSCOPY, REF DYNJ62136B; 29) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392F; 30) TOTAL LAPAROSCOPIC HYST PACK, REF DYNJ63456B; 31) ROBOTIC PACK, REF DYNJ67177D; 32) MH GYN LAPAROSCOPY PACK-LF, REF DYNJ84433B; 33) DAVINCI, REF DYNJ900963M; 34) LEX DAVINCI GYN MAJOR, REF DYNJ900965N; 35) LEX DAVINCI UROLOGY, REF DYNJ900967N; 36) COR DAVINCI, REF DYNJ901018O; 37) GYN ROBOTIC, REF DYNJ901046M; 38) DAVINCI, REF DYNJ901076L; 39) GYN ROBOTIC, REF DYNJ901796K; 40) LOU GENERAL ROBOT, REF DYNJ901844N; 41) LOU GYN LAPAROSCOPY, REF DYNJ901845L; 42) LOU GYN LAPAROSCOPY, REF DYNJ901845M; 43) LOU LITHOTOMY ROBOTIC, REF DYNJ901850N; 44) LEX GYN MINOR LAPAROSCOPY, REF DYNJ902029G; 45) GYN ROBOT, REF DYNJ902513J; 46) CHN OR DAVINCI PROSTATE, REF DYNJ902540O; 47) WMC ROBOTIC PACK - LF, REF DYNJ902583N; 48) ROBOTIC GYN & GU, REF DYNJ903459J; 49) GYN LAPAROSCOPY-LF, REF DYNJ903647I; 50) ROBOTIC UROLOGY, REF DYNJ903827F; 51) LITHOTOMY ROBOTIC CDS, REF DYNJ903948R; 52) GYN ROBOTIC, REF DYNJ903983P; 53) GYN LAP, REF DYNJ903990Q; 54) ROBOTIC-LF, REF DYNJ904254L; 55) ROBOTIC, REF DYNJ904343G; 56) ROBOT GYN, REF DYNJ904572L; 57) ROBOTIC PROSTECTOMY, REF DYNJ904597G; 58) GYN LAPAROSCOPY-LF, REF DYNJ905289G; 59) GYN ROBOTIC, REF DYNJ905730J; 60) URO ROBOTIC, REF DYNJ905746J; 61) COLON KIT, REF DYNJ905964G; 62) CHN OR ROBOTIC GYN, REF DYNJ905985O; 63) CW ROBOT, REF DYNJ906051C; 64) CW ROBOT, REF DYNJ906051D; 65) ROBOT, REF DYNJ906112B; 66) CDS ROBOTIC GYN CH, REF DYNJ906265D; 67) ROBOTIC GYN PACK, REF DYNJ906265F; 68) KIT GYN LAPAROSCOPY RFD, REF DYNJ906322F; 69) GYN LAPAROSCOPY, REF DYNJ906357C; 70) ROBOTIC COLON, REF DYNJ906379D; 71) GYN DAVINCI FOAKS, REF DYNJ906444C; 72) DAVINCI, REF DYNJ906543B; 73) CHN GYN ONCOLOGY, REF DYNJ906731K; 74) CHN ROBOTIC CYSTECTOMY, REF DYNJ907072F; 75) WMC THORACIC ROBOTIC, REF DYNJ907153C; 76) WMC THORACIC ROBOTIC, REF DYNJ907153D; 77) GYN ROBOTIC, REF DYNJ907492F; 78) KIT LAP GYN NTX, REF DYNJ907599C; 79) KIT ROBOTIC GYN, REF DYNJ907635D; 80) KIT ROBOTIC GYN, REF DYNJ907647B; 81) ROBOTIC-LF, REF DYNJ907910F; 82) ROBOTIC GENERAL, REF DYNJ907919C; 83) ROBOTIC BASE, REF DYNJ908447D; 84) ROBOTIC BASE, REF DYNJ908447F; 85) GYN, REF DYNJ908504A; 86) GU DAVINCI KIT, REF DYNJ908783F; 87) ROBOTIC PROSTATE, REF DYNJ909120D; 88) OSC ROBOT BARIATRIC, REF DYNJ909547; 89) LEX DAVINCI PROSTATE, REF DYNJ909548A; 90) GYN-URO ROBOT, REF DYNJ910137A; 91) PROSTATECTOMY ROBOTIC, REF DYNJ910991; 92) GYN ROBOTIC, REF DYNJ910995; 93) ROBOTIC, REF DYNJ9225761X; 94) DAVINCI ROBOTIC PACK-LF, REF DYNJ9288198X; 95) MB GYN ROBOTIC ADD, REF DYNJ9701096M; 96) PHS722500FCK, REF PHS722500G.
Reason for recall
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1088-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.