Data Foundry

Medical device recalls 2026

Medline medical procedure convenience kits labeled as: KIT ROBOTICS… — Class II medical device recall Z-1090-2026

Ongoing recall by Medline Industries, LP, reported 2026-01-21, distributed nationwide. Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigatio

Recall numberZ-1090-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2025-12-16
Classified2026-01-14
Reported by FDA2026-01-21
DistributedNationwide (US)
Quantity12 units
Reason classespackaging
UPC / GTIN / UDI-DI10198459156991, 40198459156992
Lot numbers25EBM426
Model numbersDYKMBNDL116G

Product

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Reason for recall

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1090-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.