Philips Azurion Systems with software versions R2 — Class II medical device recall Z-1091-2026
Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2026-01-21, distributed nationwide. Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple driv
| Recall number | Z-1091-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands |
| Recall initiated | 2025-12-12 |
| Classified | 2026-01-14 |
| Reported by FDA | 2026-01-21 |
| Distributed | Nationwide (US) |
| Countries outside the US | AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, KY, LB, LK, LT, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, UY, VN, ZA |
| Quantity | 10,466 units |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272, 00884838103276, 00884838110564, 00884838110571 |
| Model numbers | 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228, 722280, 722281, 722282 |
Product
Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12. Model Number: 722227. 4. Azurion 5 M20. Model Numbers: 722228, 722281. 5. Azurion 7 B12. Model Numbers: 722067, 722225. 6. Azurion 7 B20. Model Numbers: 722068, 722226. 7. Azurion 7 M12. Model Numbers: 722078, 722223. 8. Azurion 7 M20. Model Numbers: 722079, 722224, 722282.
Reason for recall
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.