00434811113 — Class II medical device recall Z-1092-2014
Terminated recall by Zimmer, Inc., reported 2014-03-05, distributed in IL. This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material an
| Recall number | Z-1092-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer, Inc., Warsaw, IN, United States |
| Recall initiated | 2014-01-10 |
| Classified | 2014-02-25 |
| Reported by FDA | 2014-03-05 |
| Terminated | 2014-09-11 |
| Distributed | IL |
| Quantity | 10 units |
| Reason classes | potency |
| Lot numbers | 62267250 |
| Model numbers | I4348-111-13 |
Product
00434811113, REF I4348-111-13 Zimmer Trabecular Metal Shoulder Humeral Stem 48 Degrees 11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48 Degree 11mm x 130mm, Sterile, QTY 1, Rx Product Usage: Zimmer Trabecular Metal TM Humeral Stem is a shoulder prosthetic replacement device that is indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint as well as humeral fractures. The humeral stem uses. Trabecular Metal around the proximal part of the stem for biological ingrowth.
Reason for recall
This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification. Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal. The other two zones exceeded the minimum specification.