Data Foundry

Medical device recalls 2017

Tina-Quant — Class II medical device recall Z-1092-2017

Terminated recall by Roche Diagnostics Operations, Inc., reported 2017-02-01, distributed nationwide. Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 1197285

Recall numberZ-1092-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Operations, Inc., Indianapolis, IN, United States
Recall initiated2016-02-26
Classified2017-01-24
Reported by FDA2017-02-01
Terminated2017-06-05
DistributedNationwide (US)
Quantity3,204
Lot numbers60445001
Model numbers11972855216

Product

Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High Sensitive Immunoturbidimetric assay is for the in-vitro quantitative determination of C-reactive protein (CRP) in human serum and plasma on Roche automated clinical chemistry analyzers. Highly sensitive measurement of CRP is of use for the detection and evaluation of inflammatory disorders, and associated diseases, infection and tissue injury. Measurement of CRP may also be used as an aid in the assessment of the risk of future coronary heart disease. When used as an adjunct to other laboratory evaluation methods of acute coronary syndromes, it may also be an additional independent indicator of recurrent event prognosis in patients with stable coronary disease or acute coronary syndrome

Reason for recall

Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 11972855216) shows an under-recovery of up to -25% at concentrations below 5 mg/L on the MODULAR ANALYTICS P module. this issue can lead to erroneous low CRPHS results. CRPHS is used as a marker for the prediction of cardiovascular risk. An erroneous low result could lead to an incorrect cardiovascular risk assessment and subsequent delay of further examination.