NuOss XC Sinus Reference No 509-3001 — Class II medical device recall Z-1097-2016

Terminated recall by Collagen Matrix Inc, reported 2016-03-23. During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf

Recall numberZ-1097-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCollagen Matrix Inc, Franklin Lakes, NJ, United States
Recall initiated2013-05-13
Classified2016-03-11
Reported by FDA2016-03-23
Terminated2017-06-21
Quantity347 units
Lot numbersBMCXU11H3, BMCXU11P1, BMCXU12P1

Product

NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 22 mm diameter x 12 m height) Manufacturer:- Collagen Matrix, Distributor:- ACE Surgical Supply Co, Inc, Intended use for dental surgery such as -augmentation or reconstructive treatment of alveoloar ridge -filling of infrabony periodontal defects -filling of defects after root resection, apicoectomy and cystectomy -filling of extraction sockets to enhance preservation of the alveolar ridge -elevation of maxillary sinus floor -filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) -Filling of pre-implant defects in conjunction with products intended for Guided Bone Regeneration

Reason for recall

During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the device, 3 of 13 products did not pass the recovery test, which requires that the compressed product fully expand within 1 minute of hydration time.