9180 Electrolyte Analyzer — Class II medical device recall Z-1098-2021

Terminated recall by Roche Diagnostics Operations, Inc., reported 2021-02-17, distributed nationwide. May display Calcium results on the screen with wrong arrow direction which may lead to misinterpretation of re

Recall numberZ-1098-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Operations, Inc., Indianapolis, IN, United States
Recall initiated2021-01-06
Classified2021-02-08
Reported by FDA2021-02-17
Terminated2024-02-28
DistributedNationwide (US)
Quantity1,214 units
Serial numbers12354, 12365, 12366, 12367, 12369, 12370, 12374, 12375, 12405, 12416, 12495, 12547, 12563, 12564, 12571, 12584, 12603, 12604, 12605, 12680, 12683, 12684, 12728, 12785, 12786 and 475 more
Model numbers03157334001

Product

9180 Electrolyte Analyzer - Product Usage: intended to be used for the measurement of sodium, potassium, chloride, ionized calcium and lithium in the whole blood, serum or plasma, urine, dialysate solutions, or QC materials. catalog number 03157334001 . in vitro diagnostic use.

Reason for recall

May display Calcium results on the screen with wrong arrow direction which may lead to misinterpretation of results and incorrect medical decision. If the unit for iCa++ is set to mg/dL (configured as Service Code MGL) and a measured iCa++ value is lower than the normal range, an upward arrow is shown on the display instead of a downward arrow. The display is set per default to mmol/L; therefore, the arrow indicating an alarm reflects normal ranges of iCa++ in mmol/L. If mg/dL is chosen, the arrow warnings on screen still reflect the ranges from mmol/L, rather than mg/dL. The numeric result is displayed correctly.