Data Foundry

Medical device recalls 2026

RayStation/RayPlan — Class II medical device recall Z-1104-2026

Ongoing recall by RAYSEARCH LABORATORIES AB, reported 2026-01-28. Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of

Recall numberZ-1104-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmRAYSEARCH LABORATORIES AB, Stockholm, Sweden
Recall initiated2025-11-28
Classified2026-01-16
Reported by FDA2026-01-28
Quantity22 units
UPC / GTIN / UDI-DI07350002010426, 07350002010495, 07350002010501, 07350002010570, 07350002010600
Serial numbers12.0.0.932, 12.0.3.68, 12.0.4.12, 12.1.0.1221, 12.3.0.119
Model numbers12.0.0, 12.0.3, 12.0.4, 12.1.0, 12.3.0
Expiration dates2028-03-20

Product

RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System

Reason for recall

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1104-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.