Product 7 consists of all products under product code HWC — Class II medical device recall Z-1105-2016

Terminated recall by Zimmer Manufacturing B.V., reported 2016-03-23, distributed nationwide. As a result of the insufficient sealer calibration data for product packaged in firm's Building II between Aug

Recall numberZ-1105-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer Manufacturing B.V., Mercedita, PR, United States
Recall initiated2016-01-11
Classified2016-03-12
Reported by FDA2016-03-23
Terminated2017-04-13
DistributedNationwide (US)
Quantity109
Reason classessterility
Lot numbers61800380, 61800381, 61830478, 61830479, 61903333, 62047629, 62081271, 62127760, 62133855, 62138082, 62159156, 62230449, 62230452, 62261628
Model numbers3.5MM, 47234801435

Product

Product 7 consists of all products under product code HWC, and same usage: Item no: 47234801435 3.5MM CORT SCR X 14MM 47234801635 3.5MM CORT SCR X 16MM 47234806535 3.5MM CORT SCR X 65MM 47234807035 3.5MM CORT SCR X 70MM 47234807535 3.5MM CORT SCR X 75MM 47234808035 3.5MM CORT SCR X 80MM 47234808535 3.5MM CORT SCR X 85MM 47234809035 3.5MM CORT SCR X 90MM Product Usage: A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Reason for recall

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1105-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.