Turbett Surgical Container — Class II medical device recall Z-1111-2022
Ongoing recall by TURBETT SURGICAL, LLC, reported 2022-05-25, distributed in CA, NY, PA. Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after pro
| Recall number | Z-1111-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | TURBETT SURGICAL, LLC, Victor, NY, United States |
| Recall initiated | 2022-04-01 |
| Classified | 2022-05-18 |
| Reported by FDA | 2022-05-25 |
| Distributed | CA, NY, PA |
| Quantity | 232 units |
| Reason classes | sterility |
| UPC / GTIN / UDI-DI | 00859464007113 |
| Lot numbers | 100001, 100004, 100036, 100065, 100083, 100099, 100125, 100130, 100150, 100154, 100173, 100181, 160522-1, 2016071000, 2016071001, 2016071002, 2016071003, 2016071005, 2016071006, 2016071007, 2016071008, 2016071009, 2016071010, 2016071011, 2016071012, 2016081000, 2016081001, 2016081002, 2016081003, 2016081004, 2016081005, 2016081006, 2016081007, 2016081008, 2016081009, 2016091000, 2016091001, 2016091002, 2016091003, 2016091004 and 203 more |
Product
Turbett Surgical Container, TS1500
Reason for recall
Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after prolonged use.