Turbett Surgical Container — Class II medical device recall Z-1111-2022

Ongoing recall by TURBETT SURGICAL, LLC, reported 2022-05-25, distributed in CA, NY, PA. Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after pro

Recall numberZ-1111-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmTURBETT SURGICAL, LLC, Victor, NY, United States
Recall initiated2022-04-01
Classified2022-05-18
Reported by FDA2022-05-25
DistributedCA, NY, PA
Quantity232 units
Reason classessterility
UPC / GTIN / UDI-DI00859464007113
Lot numbers100001, 100004, 100036, 100065, 100083, 100099, 100125, 100130, 100150, 100154, 100173, 100181, 160522-1, 2016071000, 2016071001, 2016071002, 2016071003, 2016071005, 2016071006, 2016071007, 2016071008, 2016071009, 2016071010, 2016071011, 2016071012, 2016081000, 2016081001, 2016081002, 2016081003, 2016081004, 2016081005, 2016081006, 2016081007, 2016081008, 2016081009, 2016091000, 2016091001, 2016091002, 2016091003, 2016091004 and 203 more

Product

Turbett Surgical Container, TS1500

Reason for recall

Eight units manufactured before June 2019 have exhibited weld failures breaching the sterile barrier after prolonged use.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1111-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.