Data Foundry

Medical device recalls 2015

Terumo HX2 Temperature Management System provides temperature control… — Class II medical device recall Z-1116-2015

Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2015-02-25, distributed nationwide. Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and sho

Recall numberZ-1116-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTerumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States
Recall initiated2015-02-06
Classified2015-02-13
Reported by FDA2015-02-25
Terminated2016-10-17
DistributedNationwide (US)
Countries outside the USJP
Quantity103 units
Reason classesdevice malfunction
Serial numbers0001, 00016, 0002, 1001, 1002, 1042, 1116, HX2, PT0003, PT0004
Model numbers809810, 809810TK, 812231, 812240

Product

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.

Reason for recall

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1116-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.