Edwards Lifesciences — Class I medical device recall Z-1117-2019
Terminated recall by Edwards Lifesciences, LLC, reported 2019-05-01, distributed nationwide. Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of balloon fragmentat
| Recall number | Z-1117-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Edwards Lifesciences, LLC, Irvine, CA, United States |
| Recall initiated | 2019-03-28 |
| Classified | 2019-04-25 |
| Reported by FDA | 2019-05-01 |
| Terminated | 2023-11-01 |
| Distributed | Nationwide (US) |
| Quantity | 1,564 |
| Reason classes | device malfunction |
| Model numbers | 830515F |
Product
Edwards Lifesciences, Miller Balloon Atrioseptostomy Catheter, Sterile EO, Rx Only, REF: 830515F Product Usage: The Miller balloon atrioseptostomy catheter is designed for enlarging interatrial openings. In most pediatric cardiology centers, balloon atrioseptostomy is an accepted technique for palliation of several congenital cardiac defects. Description of Use Balloon atrioseptostomy can be performed in conjunction with diagnostic cardiac catheterization or under echocardiographic guidance and has been carried out after the diagnosis of several congenital cardiac defects including but not limited to: transposition of the great arteries, total anomalous pulmonary venous drainage without pulmonary obstruction, tricuspid atresia, and pulmonary atresia with intact ventricular septum.
Reason for recall
Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of balloon fragmentation or detachment.