Data Foundry

Medical device recalls 2026

ICU Medical Plum Duo Precision IV Pump — Class II medical device recall Z-1119-2026

Ongoing recall by ICU Medical, Inc., reported 2026-01-28, distributed nationwide. Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream l

Recall numberZ-1119-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmICU Medical, Inc., Lake Forest, IL, United States
Recall initiated2025-12-15
Classified2026-01-16
Reported by FDA2026-01-28
DistributedNationwide (US)
Countries outside the USPH

Product

ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)

Reason for recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1119-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.