Product 22 consists of all product under product code: JDI and same… — Class II medical device recall Z-1120-2016
Terminated recall by Zimmer Manufacturing B.V., reported 2016-03-23, distributed nationwide. As a result of the insufficient sealer calibration data for product packaged in firm's Building II between Aug
| Recall number | Z-1120-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer Manufacturing B.V., Mercedita, PR, United States |
| Recall initiated | 2016-01-11 |
| Classified | 2016-03-12 |
| Reported by FDA | 2016-03-23 |
| Terminated | 2017-04-13 |
| Distributed | Nationwide (US) |
| Quantity | 16 |
| Reason classes | sterility |
| Lot numbers | 61907682, 62030932, 62038367, 62106796, 62106797, 62148557, 62161479, 62206538, 62240742, 62240743, 62268577, 62303507, 62344683, 62352823 |
| Model numbers | 12/14, 811400218CPT |
Product
Product 22 consists of all product under product code: JDI and same usage: Item no: 811400218 CPT 12/14 COCR REVISION S 811400318 CPT 12/14 COCR REVISION S 811400420 CPT 12/14 COCR REVISION S 811400423 CPT 12/14 COCR REVISION S 811400426 CPT 12/14 COCR REVISION S 811400010 CPT 12/14 COCR SIZE 0 EXT 811400000 CPT 12/14 COCR SIZE 0 STD 811400110 CPT 12/14 COCR SIZE 1 EXT 811400100 CPT 12/14 COCR SIZE 1 STD 811400210 CPT 12/14 COCR SIZE 2 EXT 811400200 CPT 12/14 COCR SIZE 2 STD 811400230 CPT 12/14 COCR SIZE 2 XEX 811400310 CPT 12/14 COCR SIZE 3 EXT 811400300 CPT 12/14 COCR SIZE 3 STD 811400330 CPT 12/14 COCR SIZE 3 XEX 811400410 CPT 12/14 COCR SIZE 4 EXT 811400400 CPT 12/14 COCR SIZE 4 STD 811400430 CPT 12/14 COCR SIZE 4 XEX 811400510 CPT 12/14 COCR SIZE 5 EXT 811400500 CPT 12/14 COCR SIZE 5 STD 811400530 CPT 12/14 COCR SIZE 5 XEX For use in total or hemi hip arthroplasty
Reason for recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1120-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.