COBAS INTEGRA 400 plus analyzer used in the following assays:… — Class II medical device recall Z-1124-2019
Terminated recall by Roche Diagnostics Corporation, reported 2019-04-17, distributed nationwide. COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode
| Recall number | Z-1124-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Corporation, Indianapolis, IN, United States |
| Recall initiated | 2018-12-14 |
| Classified | 2019-04-10 |
| Reported by FDA | 2019-04-17 |
| Terminated | 2020-09-04 |
| Distributed | Nationwide (US) |
| Quantity | 656 units |
| Model numbers | 03032566122, 03263991190, 04469658190, 20767107322 |
Product
COBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin Gen.2 Catalog Number 04469658190 c)Creatinine plus ver.2 Catalog Number 03263991190 d) riglycerides Catalog Number :20767107322 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems
Reason for recall
COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1124-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.