Product 27 consists of all product under product code: HSB and same… — Class II medical device recall Z-1125-2016
Terminated recall by Zimmer Manufacturing B.V., reported 2016-03-23, distributed nationwide. As a result of the insufficient sealer calibration data for product packaged in firm's Building II between Aug
| Recall number | Z-1125-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer Manufacturing B.V., Mercedita, PR, United States |
| Recall initiated | 2016-01-11 |
| Classified | 2016-03-12 |
| Reported by FDA | 2016-03-23 |
| Terminated | 2017-04-13 |
| Distributed | Nationwide (US) |
| Quantity | 945 |
| Reason classes | sterility |
| Lot numbers | 61622044, 61634097, 61635839, 61690420, 61690421, 61690423, 61695534, 61700084, 61714629, 61714630, 61777525, 61777526, 61777527, 61777528, 61777529, 61777530, 61777531, 61777535, 61777537, 61782870, 61782872, 61782874, 61782876, 61782878, 61782879, 61782881, 61782882, 61782883, 61782884, 61782888, 61782889, 61782891, 61793329, 61801382, 61801383, 61801384, 61801385, 61801389, 61804353, 61804354 and 460 more |
| Model numbers | 234703104 |
Product
Product 27 consists of all product under product code: HSB and same usage: Item no: 234703104 DISTAL MEDIAL FEMORAL CON 234703206 DISTAL MEDIAL FEMORAL CON 234703208 DISTAL MEDIAL FEMORAL CON 234800504 DISTAL POSTERIOR/LATERAL 234800506 DISTAL POSTERIOR/LATERAL 234800508 DISTAL POSTERIOR/LATERAL 234800510 DISTAL POSTERIOR/LATERAL 234800604 DISTAL POSTERIOR/LATERAL 234800606 DISTAL POSTERIOR/LATERAL 234800608 DISTAL POSTERIOR/LATERAL 234800610 DISTAL POSTERIOR/LATERAL 234800614 DISTAL POSTERIOR/LATERAL Product Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Reason for recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.