REPAIR KIT — Class II medical device recall Z-1127-2018

Terminated recall by Medical Components, Inc dba MedComp, reported 2018-03-28, distributed nationwide. There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubin

Recall numberZ-1127-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedical Components, Inc dba MedComp, Harleysville, PA, United States
Recall initiated2018-02-23
Classified2018-03-19
Reported by FDA2018-03-28
Terminated2020-01-10
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, CA, DE, EC, ES, FR, GB, GR, IE, IL, IT, NL, SE, TR, ZA
Quantity1,575 kits
UPC / GTIN / UDI-DI00884908039856
Lot numbersMCCA800, MCCC310, MCCC320, MCCL340, MCDD250, MCDD640, MCDP460, MCFA160, MCFC040, MCFH990, MCFS390, MCFW690, MKAB730
Model numbersRPK-01

Product

REPAIR KIT, REF RPK-01, UDI 884908039856 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.

Reason for recall

There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.