SR Instruments FG3105 Patient Scales Used on various Joerns… — Class II medical device recall Z-1128-2013
Terminated recall by Sr Instruments Inc, reported 2013-04-24, distributed nationwide. SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they rece
| Recall number | Z-1128-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Sr Instruments Inc, Tonawanda, NY, United States |
| Recall initiated | 2012-12-13 |
| Classified | 2013-04-16 |
| Reported by FDA | 2013-04-24 |
| Terminated | 2015-03-12 |
| Distributed | Nationwide (US) |
| Quantity | 806 patient |
| Serial numbers | 1005 |
| Model numbers | FG3105 |
Product
SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.
Reason for recall
SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.