Data Foundry

Medical device recalls 2013

SR Instruments FG3105 Patient Scales Used on various Joerns… — Class II medical device recall Z-1128-2013

Terminated recall by Sr Instruments Inc, reported 2013-04-24, distributed nationwide. SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they rece

Recall numberZ-1128-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSr Instruments Inc, Tonawanda, NY, United States
Recall initiated2012-12-13
Classified2013-04-16
Reported by FDA2013-04-24
Terminated2015-03-12
DistributedNationwide (US)
Quantity806 patient
Serial numbers1005
Model numbersFG3105

Product

SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.

Reason for recall

SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.