Data Foundry

Medical device recalls 2026

The LDL-Cholesterol assay is comprised of two distinct phases — Class II medical device recall Z-1135-2026

Ongoing recall by Beckman Coulter Ireland, Inc., reported 2026-01-28, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Ins

Recall numberZ-1135-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Ireland, Inc., O'Callaghan'S Mills, Ireland
Recall initiated2025-12-08
Classified2026-01-21
Reported by FDA2026-01-28
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY
Countries outside the USAR, BR, CA, CN, CR, DE, EC, GB, IE, KR, MX, PA, PE, SG, SV, ZA
Quantity34,451
UPC / GTIN / UDI-DI15099590010928, 15099590020361

Product

The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.

Reason for recall

Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1135-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.