The LDL-Cholesterol assay is comprised of two distinct phases — Class II medical device recall Z-1135-2026
Ongoing recall by Beckman Coulter Ireland, Inc., reported 2026-01-28, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Ins
| Recall number | Z-1135-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter Ireland, Inc., O'Callaghan'S Mills, Ireland |
| Recall initiated | 2025-12-08 |
| Classified | 2026-01-21 |
| Reported by FDA | 2026-01-28 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY |
| Countries outside the US | AR, BR, CA, CN, CR, DE, EC, GB, IE, KR, MX, PA, PE, SG, SV, ZA |
| Quantity | 34,451 |
| UPC / GTIN / UDI-DI | 15099590010928, 15099590020361 |
Product
The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.
Reason for recall
Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).