Product 40 consists of all product under product code: HSB and same… — Class II medical device recall Z-1138-2016
Terminated recall by Zimmer Manufacturing B.V., reported 2016-03-23, distributed nationwide. As a result of the insufficient sealer calibration data for product packaged in firm's Building II between Aug
| Recall number | Z-1138-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer Manufacturing B.V., Mercedita, PR, United States |
| Recall initiated | 2016-01-11 |
| Classified | 2016-03-12 |
| Reported by FDA | 2016-03-23 |
| Terminated | 2017-04-13 |
| Distributed | Nationwide (US) |
| Quantity | 7 |
| Reason classes | sterility |
| Lot numbers | 62042715, 62042716, 62208146, 62208147, 62208148 |
| Model numbers | 225706565 |
Product
Product 40 consists of all product under product code: HSB and same usage: Item no: 225706565 ITST ANTI-ROTATION SCREW 225707065 ITST ANTI-ROTATION SCREW 225708065 ITST ANTI-ROTATION SCREW 225708565 ITST ANTI-ROTATION SCREW 225709065 ITST ANTI-ROTATION SCREW Product Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Reason for recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1138-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.