Terminated recall by Zimmer Manufacturing B.V., reported 2016-03-23, distributed nationwide. As a result of the insufficient sealer calibration data for product packaged in firm's Building II between Aug

Recall numberZ-1138-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer Manufacturing B.V., Mercedita, PR, United States
Recall initiated2016-01-11
Classified2016-03-12
Reported by FDA2016-03-23
Terminated2017-04-13
DistributedNationwide (US)
Quantity7
Reason classessterility
Lot numbers62042715, 62042716, 62208146, 62208147, 62208148
Model numbers225706565

Product

Product 40 consists of all product under product code: HSB and same usage: Item no: 225706565 ITST ANTI-ROTATION SCREW 225707065 ITST ANTI-ROTATION SCREW 225708065 ITST ANTI-ROTATION SCREW 225708565 ITST ANTI-ROTATION SCREW 225709065 ITST ANTI-ROTATION SCREW Product Usage: An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Reason for recall

As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.