Data Foundry

Medical device recalls 2018

BD Vacutainer¿ Lithium Heparin Green Top Tube — Class I medical device recall Z-1142-2018

Terminated recall by Becton Dickinson & Company, reported 2018-03-28, distributed in AL, AR, AZ, CA, CO, CT, FL, GA…. BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cau

Recall numberZ-1142-2018
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBecton Dickinson & Company, Franklin Lakes, NJ, United States
Recall initiated2018-03-22
Classified2018-03-22
Reported by FDA2018-03-28
Terminated2022-07-25
DistributedAL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV
Quantity117,816,603
Reason classespotency
UPC / GTIN / UDI-DI50382903666640, 50382903666671, 50382903678803, 50382903678841, 50382903678865, 50382903680400
Model numbers366664, 366667, 367880, 367884, 367886, 368040

Product

BD Vacutainer¿ Lithium Heparin Green Top Tube; Catalog Numbers: 1. 366664 (UDI: 50382903666640); 2. 367884 (UDI: 50382903678841); 3. 367886 (UDI: 50382903678865); 4. 368040 (UDI: 50382903680400); 5. 366667 (UDI: 50382903666671); 6. 367880 (UDI: 50382903678803)

Reason for recall

BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1142-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.