Trilogy Ventilators The Respironics Trilogy 100 system provides… — Class I medical device recall Z-1143-2014
Terminated recall by Phillips Respironics, Inc., reported 2014-03-12, distributed nationwide. Internal testing revealed a potentially defective component on the power management board that could affect th
| Recall number | Z-1143-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Phillips Respironics, Inc., Murrysville, PA, United States |
| Recall initiated | 2014-02-11 |
| Classified | 2014-03-06 |
| Reported by FDA | 2014-03-12 |
| Terminated | 2015-04-17 |
| Distributed | Nationwide (US) |
| Countries outside the US | CN, FR, IN, IT, KR, KW, SG |
| Quantity | 572 |
Product
Trilogy Ventilators The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilogy 100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator. Trilogy 200 Intended Use: The Respironics Trilogy 200 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilogy 200 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. Trilogy 202 Intended Use: The Respironics Trilogy 202 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation with or without air/oxygen blending. Trilogy 202 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in hospitals and institutions, and for portable applications such as wheelchairs and gurneys only when in an institutional setting. It may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.
Reason for recall
Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.