ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 — Class II medical device recall Z-1143-2023
Ongoing recall by Siemens Healthcare Diagnostics, Inc., reported 2023-03-01, distributed nationwide. Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hep
| Recall number | Z-1143-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., East Walpole, MA, United States |
| Recall initiated | 2022-12-22 |
| Classified | 2023-02-17 |
| Reported by FDA | 2023-03-01 |
| Distributed | Nationwide (US) |
| Quantity | 158 units |
| Reason classes | potency |
| UPC / GTIN / UDI-DI | 00630414167923 |
| Lot numbers | B34847, B34848, B35113 |
Product
ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection of antibodies to the e antigen of the hepatitis B virus (HBV) in human pediatric (2-21 years old) and adult serum, EDTA plasma, or lithium heparin. Siemens Material Number (SMN): 10720831
Reason for recall
Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive when HBeAg testing follows aHBe2 testing, regardless of whether both tests are performed on the same sample