Data Foundry

Medical device recalls 2015

Spacelabs Healthcare qube Compact Monitor — Class II medical device recall Z-1145-2015

Terminated recall by Spacelabs Healthcare Inc, reported 2015-03-04. Spacelabs Healthcare qube Bedside Monitors, Model 91390, are recalled because the firm has received multiple r

Recall numberZ-1145-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSpacelabs Healthcare Inc, Snoqualmie, WA, United States
Recall initiated2015-01-28
Classified2015-02-24
Reported by FDA2015-03-04
Terminated2015-12-30
Model numbers91390

Product

Spacelabs Healthcare qube Compact Monitor, Model 91390. The Spacelabs Healthcare qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output.

Reason for recall

Spacelabs Healthcare qube Bedside Monitors, Model 91390, are recalled because the firm has received multiple reports of qube monitors failing to boot up or returning to factory default configuration settings following power on or reset.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1145-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.