The Medtronic — Class II medical device recall Z-1147-2015
Terminated recall by Medtronic Neuromodulation, reported 2015-03-04, distributed nationwide. Current labeling indicates that cycling improves device longevity in every programming scenario. However, enab
| Recall number | Z-1147-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Neuromodulation, Minneapolis, MN, United States |
| Recall initiated | 2015-01-09 |
| Classified | 2015-02-24 |
| Reported by FDA | 2015-03-04 |
| Terminated | 2017-09-28 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, MX |
| Reason classes | labeling |
| Model numbers | 3023, 3058, 3116, 7425 |
Product
The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or muscles when connected to an extension and/or lead. The Medtronic InterStim II Neurostimulator, Model 3058, and Medtronic InterStim Neurostimulator, Model 3023, are parts of neurostimulation systems for InterStim Therapy. The neurostimulators are multiprogrammable devices that accommodates a lead through which a stimulation program is delivered. The Medtronic Enterra Neurostimulator, Model 3116, is part of an Enterra Therapy system for gastric electrical stimulation (GES). The neurostimulator is a multiprogrammable device that accommodates a lead through which a stimulation program is delivered.
Reason for recall
Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1147-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.