Data Foundry

Medical device recalls 2014

GE Healthcare — Class II medical device recall Z-1148-2014

Terminated recall by GE Healthcare, LLC, reported 2014-03-12, distributed nationwide. When performing a secondary capture generated from GSI Viewer, the header information is correct on the images

Recall numberZ-1148-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2011-03-25
Classified2014-03-06
Reported by FDA2014-03-12
Terminated2014-03-13
DistributedNationwide (US)
Countries outside the USAU, BE, BR, CA, CH, CN, DE, DK, FI, FR, GB, GR, HK, ID, IL, IN, IT, JP, KR, KW, LV, NZ, PL, SA, SE, SG, TW, ZA
Serial numbers00000018579HP0, 000000G9337001, 000000G9337002, 000000G9337003, 000000G9337004, 000000G9337005, 000000G9337006, 000000G9337008, 000000G9337009, 000000G9337010, 000000G9337012, 000000G9337013, 000000G9337014, 000000G9337015, 000000G9337017, 000000G9337018, 000000G9337019, 000000G9337020, 000000L9343001, 000000L9343002, 000000L9343003, 000000L9343004, 000000L9343005, 000000L9343006, 000000L9343007 and 258 more

Product

GE Healthcare, Model Number/ Description: 5342114 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-2 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-3 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-4 GSI Viewer 1.11 on AW VolumeShare 4 (AW 4.5); 5401757 GSI Viewer 1.20 on AW VolumeShare 3 (AW4.4). Computed Tomography X-Ray system

Reason for recall

When performing a secondary capture generated from GSI Viewer, the header information is correct on the images. However, when changing the ROI (region of interest) and zooming in, the DICOM header does not change to reflect the new image, thus the line measurements may be inaccurate.