Epidural tray — Class II medical device recall Z-1150-2024
Ongoing recall by B. Braun Medical, Inc., reported 2024-02-28, distributed nationwide. Kits should contain Filter Straws with Standard Luer Connections however, the kits contain Filter Straws with
| Recall number | Z-1150-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | B. Braun Medical, Inc., Allentown, PA, United States |
| Recall initiated | 2024-01-22 |
| Classified | 2024-02-16 |
| Reported by FDA | 2024-02-28 |
| Distributed | Nationwide (US) |
| Quantity | 3,890 units |
| UPC / GTIN / UDI-DI | 04046964178689 |
| Lot numbers | 0061855122 |
| Expiration dates | 2024-10-20 |
Product
Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of devices intended for the administration of an analgesic or anesthetic agent to the epidural space for the management of pain. It includes the pharmaceutical anesthetic agent and a needle, with additional devices utilized for the procedure (e.g., syringe, dressings), and may include a catheter (i.e., may be intended for single- or continuous-administration). It is not intended for intrathecal anesthesia. Product Code: 332248
Reason for recall
Kits should contain Filter Straws with Standard Luer Connections however, the kits contain Filter Straws with Neuraxial connectors preventing interface with the other components of the kit.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1150-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.